Enameled wire tester

High Voltage Continuity of Covering Tester

High Voltage Continuity of Covering Tester

Instrument Description

High voltage

- A voltage generator must provide the electrode with a continuous filtered voltage without transients. The test voltages using an open circuit must be adjustable from 350  to 3000Vdc+-5%.The permanent short circuit current must be 25±5μA  at any voltage. A 50 Mohm resistance  applied to the high voltage electrode should not cause a voltage drop  more than  75% irrespective of the voltage used; the specimen length shall be 30 meters. An appropriate circuit  must evaluate whether the wire carries a current which is greater than that specified in the tables; a counter will display the number of faults.


- Suitable for diameters range 0.05 up to 1.6 mm (44 -14 AWG).

- Test voltages adjustable in 8 steps: 350V, 500V, 750V, 1KV, 1.5KV, 2KV, 2.5KV, 3KV

- Test current according to standard specifications.

- Four digit meter counter with pre-selectable test length,resolution 0.1 meter.

- Four digit fault counter with sign-calling maximum admissible faults.

- Winding wheel for easy wire removal, asynchronous motor with worm gearbox, adjustable speed from 3 up to 30 m/1with inverter driver.

- Supplied with electrodes according to IEC 60851 or carbon fibre electrode according to NEMA.


- Dimensions: W500×W600×D620mm

- Power supply: 230AV 50/60Hz one phase 250VA

- Weight: 38KG


Options

- GS Stretching unit: 2-4-6-8-10-12-14-16-18-20%.

- E Electrodes according IEC or NEMA.

- V Different test voltages